Multiport epidural catheter without port and incomplete marking


Department of Anaesthesiology & Critical Care, Jawaharlal Institute of Post-Graduate Medical Education and Research, Pondicherry, India

Date of Web Publication12-Aug-2010

Correspondence Address:
Sandeep Kumar Mishra
29-IV-Cross, Indira Nagar Extention, Gorimedu, Pondicherry-6
India
Login to access the Email id


DOI: 10.4103/0019-5049.68387

PMID: 20882192

Get Permissions


How to cite this article:
Gupta SL, Mishra SK, Elakkumanan LB, Sudeep K. Multiport epidural catheter without port and incomplete marking. Indian J Anaesth 2010;54:359-60

How to cite this URL:
Gupta SL, Mishra SK, Elakkumanan LB, Sudeep K. Multiport epidural catheter without port and incomplete marking. Indian J Anaesth [serial online] 2010 [cited 2013 Jun 18];54:359-60. Available from: http://www.ijaweb.org/text.asp?2010/54/4/359/68387

Sir,

We would like to report a faulty epidural catheter responsible for failure to inject local anaesthestics through it. A case of 50-year-old woman in advanced carcinoma of the cervix posted for Wertheim's hysterectomy, planned under combined epidural-general anaesthesia. Monitors were connected after securing an intravenous access. Localisation of the epidural space was achieved via the L2-L3 interspace with the patient in the lateral position. A perifix continuous Anaesthesia kit was used. An 18-gauge Tuohy epidural needle was placed in the epidural space without difficulty using the loss-of resistance technique of air. Epidural catheter was inserted through the epidural needle. After inserting few centimeters, it was noticed that markings of the epidural catheter were absent up to 14 cm. (just after three marking i.e., at 15 cm). Catheter is inserted up to 15 cm marking with approximately 4 cm inside epidural space. Injection of the test dose via the catheter was impossible. Incremental withdrawal of the catheter did not correct this situation. The epidural catheter was eventually withdrawn completely. It was not possible with subsequent attempts to flush the catheter. Close inspection of the epidural catheter assembly unit showed that a complete absence of multiple port at the distal end (helical "eyes") with incomplete markings [Figure 1]. Pre-insertion checking of the catheter and flushing the catheter ("injection test) would have averted this incident. Difficult or impossible injection via the epidural catheter can be a result commonly from mechanical obstruction of the epidural catheter at various levels, like accidental kinking, knotting and malposition of the catheter, occasional manufacturing defects of the catheter [1],[2] (defect of the screw-cap connector) can lead to this problem. Before inserting catheter, "injection test" should be performed whenever feasible, not only for epidural catheter but also for other catheter like central venous catheter.
Figure 1: Defective epidural catheter

Click here to view


 
   References Top

1.Chandhok D, Vijayakumar E. An unusual case of epidural catheter obstruction. Anaesthesiology 1999;91:895-6.  Back to cited text no. 1      
2.Husemeyer RP. A defective epidural cannula. Anaesthesia 1980;35:922.  Back to cited text no. 2      


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed1058    
    Printed49    
    Emailed0    
    PDF Downloaded100    
    Comments [Add]    

Recommend this journal