Unusual bulging (ballooning) of the laryngeal mask airway cuff causing secondary loss of airway


Department of Anaesthesiology and Critical Care, Al Jahra Hospital, Ministry of Health, Kuwait

Date of Web Publication17-May-2012

Correspondence Address:
Surjya Prasad Upadhyay
Registrar, Anaesthesiology and Critical Care, Al Jahra Hospital, Ministry of Health
Kuwait
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.96328

Rights and Permissions

How to cite this article:
Upadhyay SP, Mallick PN, Jagia M. Unusual bulging (ballooning) of the laryngeal mask airway cuff causing secondary loss of airway. Indian J Anaesth 2012;56:193-4

How to cite this URL:
Upadhyay SP, Mallick PN, Jagia M. Unusual bulging (ballooning) of the laryngeal mask airway cuff causing secondary loss of airway. Indian J Anaesth [serial online] 2012 [cited 2019 Dec 14];56:193-4. Available from: http://www.ijaweb.org/text.asp?2012/56/2/193/96328


   Introduction Top


Laryngeal mask airway (LMA) has been universally used either as primary or rescue airway devices. Although it is relatively safe and effective as a supraglottic airway device, there are few reports of its complications mostly related to its cuff volume and pressure or its malposition. Complication ranges from sore throat to nerve palsies (lingual, recurrent laryngeal, inferior alveolar, hypoglossal). [1],[2],[3],[4],[5] We encountered an unusual bulging of the disposable LMA cuff after correct positioning in hypopharynx and normal inflation with recommended volume of air. The unusual bulging (ballooning) of the cuff occurred 30 min after insertion, causing loss of ventilation and significant desaturation requiring change of LMA.


   Case Report Top


A 30-year-old male patient weighing 75 kg was scheduled for K wire fixation for bimalleolar fracture of right ankle. General anaesthesia with spontaneous ventilation through LMA was planned. After connecting standard monitors, the patient was induced with Remifentanil 50 μg over 1 min and propofol 150 mg. After checking the cuff leak and cuff deformity, Size 4 disposable LMA was inserted as per the standard technique with the index finger guidance and the cuff was inflated with 30 mL of air. LMA was connected to the ventilator tubing and patient was allowed to breathe spontaneously after initial manual assisted ventilation. Anaesthesia was maintained with oxygen 40% in air with sevoflurane (1-3%) and infusion of Remifentanil at a rate of 300-600 μg/h. Patient was breathing smoothly with good tidal exchange for the first 30 min, the patient started desaturating and the ventilator showed progressive loss of tidal exchange. All the external tubing and connections were checked and found to be intact. On manual ventilation, significant air leak from the LMA was diagnosed. LMA was taken out and replaced with another one [Figure 1] and [Figure 2].
Figure 1: Side view

Click here to view
Figure 2: Top view

Click here to view



   Discussion Top


The LMA used in this case was a new disposable unit with silicon cuff. The cuff of the LMA was inflated with air and tested before insertion, and nitrous oxide was not used intraoperatively. Therefore, the cause of this unusual ballooning (bulging one portion) was difficult to explain. Temperature in the hypopharynx might have caused expansion of the cuff volume and stretching of the silicon cuff, causing unusual bulging. We inflated the cuff with the recommended volume (30 mL for no. 4 LMA) and the patient was in a supine position. The surgical manipulation causing head and neck movement was negligible as the surgical field was far away from the body. There might be a possibility of manufacturing defect in the form of uneven thickness of silicon cuff that might have given way under normal volume and pressure. It would have been more interesting if we could have measured the cuff pressure.

The disposable LMA cuff is made of soft silicon designed to provide low pressure [6] and bulge with small increase in intracuff volume in a relatively warm atmosphere of hypopharynx, which might have lead to secondary loss of airway. In our opinion, this potential condition should be kept in mind while using any LMA that might give rise to unexpected airway complications.

 
   References Top

1.Twigg S, Brown JM, William R. Swelling and cyanosis of the tongue associated with use of a laryngeal mask airway. Anaesth Intensive Care 2000;28:449-50.  Back to cited text no. 1
    
2.Nagai K, Sakuramoto C, Goto F. Unilateral hypoglossal nerve paralysis following the use of the laryngeal mask airway. Anaesthesia 1994;49:603-4.  Back to cited text no. 2
[PUBMED]    
3.Lowinger D, Benjamin B, Gadd L. Recurrent laryngeal nerve injury caused by a laryngeal mask airway. Anaesth Intensive Care 1999;27:202-5.  Back to cited text no. 3
[PUBMED]    
4.Brimacombe J, Holyoake L, Keller C, Barry J, Mecklem D, Blinco A, et al. Emergence characteristics and postoperative laryngopharyngeal morbidity with the laryngeal mask airway: A comparison of high versus low cuff volume. Anaesthesia 2000;55:338-43.  Back to cited text no. 4
[PUBMED]  [FULLTEXT]  
5.Hanumanthaiah D, Masud S, Ranganath A. Inferior alveolar nerve injury with laryngeal mask airway: A case report. J Med Case Reports 2011;5:122.  Back to cited text no. 5
    
6.Van Zundert AA, Fonck K, Al-Shaikh B, Mortier EP. Comparison of cuff-pressure changes in LMA-Classic and the new Soft Seal laryngeal masks during nitrous oxide anaesthesia in spontaneous breathing patients. Eur J Anaesthesiol 2004;21:547-52.  Back to cited text no. 6
[PUBMED]    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   Introduction
   Case Report
   Discussion
    References
    Article Figures

 Article Access Statistics
    Viewed1329    
    Printed59    
    Emailed0    
    PDF Downloaded298    
    Comments [Add]    

Recommend this journal