REVIEW ARTICLE
Year : 2013  |  Volume : 57  |  Issue : 5  |  Page : 533-540  

Anaesthesia machine: Checklist, hazards, scavenging


Department of Anaesthesiology, Kasturba Medical College, Manipal University, Manipal, Karnataka, India

Date of Web Publication22-Oct-2013

Correspondence Address:
Umesh Goneppanavar
Department of Anaesthesiology, Kasturba Medical College, Manipal University, Manipal - 576 104, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.120151

Rights and Permissions
 

From a simple pneumatic device of the early 20 th century, the anaesthesia machine has evolved to incorporate various mechanical, electrical and electronic components to be more appropriately called anaesthesia workstation. Modern machines have overcome many drawbacks associated with the older machines. However, addition of several mechanical, electronic and electric components has contributed to recurrence of some of the older problems such as leak or obstruction attributable to newer gadgets and development of newer problems. No single checklist can satisfactorily test the integrity and safety of all existing anaesthesia machines due to their complex nature as well as variations in design among manufacturers. Human factors have contributed to greater complications than machine faults. Therefore, better understanding of the basics of anaesthesia machine and checking each component of the machine for proper functioning prior to use is essential to minimise these hazards. Clear documentation of regular and appropriate servicing of the anaesthesia machine, its components and their satisfactory functioning following servicing and repair is also equally important. Trace anaesthetic gases polluting the theatre atmosphere can have several adverse effects on the health of theatre personnel. Therefore, safe disposal of these gases away from the workplace with efficiently functioning scavenging system is necessary. Other ways of minimising atmospheric pollution such as gas delivery equipment with negligible leaks, low flow anaesthesia, minimal leak around the airway equipment (facemask, tracheal tube, laryngeal mask airway, etc.) more than 15 air changes/hour and total intravenous anaesthesia should also be considered.

Keywords: Anaesthesia machine, anaesthesia workstation, checklist, hazards, scavenging


How to cite this article:
Goneppanavar U, Prabhu M. Anaesthesia machine: Checklist, hazards, scavenging. Indian J Anaesth 2013;57:533-40

How to cite this URL:
Goneppanavar U, Prabhu M. Anaesthesia machine: Checklist, hazards, scavenging. Indian J Anaesth [serial online] 2013 [cited 2017 Feb 19];57:533-40. Available from: http://www.ijaweb.org/text.asp?2013/57/5/533/120151


   Introduction Top


Anaesthesia machine is designed to deliver O 2 along with other anaesthetic gases including volatile anaesthetic vapours in specified concentrations to patients with the help of breathing circuits. From a simple pneumatic device of the early 20 th century, the anaesthesia machine has evolved to incorporate various mechanical, electrical and electronic components to be more appropriately called anaesthesia workstation. [1] The focus of this article is to format a concise anaesthesia machine check protocol, discuss the hazards related to modern anaesthesia machines and detail the requirements for an efficient scavenging system.


   Anaesthesia Machine Check Protocol Top


Checking each component of anaesthesia machine for appropriate functioning prior to use is essential to ensure patient safety. However, a single checklist cannot satisfactorily test the integrity and safety of all existing anaesthesia machines due to their complex nature as well as variations in design among manufacturers. An in-depth and elaborate anaesthesia machine check should be done following servicing of the anaesthesia machine. Further, machine check should be done daily prior to first use [Figure 1]. This should be user friendly and less time consuming while also ensuring satisfactory check of all components of the machine. Minor check procedure should be followed between anaesthetic conducts.
Figure 1: A concise anaesthesia machine checklist for daily use (content in bold indicates minor check procedure that should be followed between anaesthetic conducts). The checklist should be modified to suit the type of anaesthesia machine / workstation available at individual location

Click here to view


First priority is to ensure the machine is placed in a safe area and its electrical wiring safely secured. Despite advanced technology, a remote but life-threatening possibility of intraoperative machine malfunction exists. Therefore, presence of a functioning self-inflating bag appropriate for the patient's age and alternate O 2 source should be ensured. [2],[3] Modern anaesthesia machine check includes the cylinders, pipelines, machine proper (both intermediate and low pressure systems and components thereof), vapourisers, breathing circuits, monitors, integrated ventilator, suction apparatus and the scavenging system. However, several modern machines perform self-check as soon as the master switch is turned on. Unnecessary repetition and missing of some component check can be prevented with adequate understanding of the components tested during self-check.

Several international guidelines are available for anaesthesia machine check. [2],[3],[4],[5] The following protocol was developed based on the existing literature and our department practices, which involves the checking for the pneumatic, electrical, electronic and other components of the machine in a systematic manner.

Checking the pneumatic components of anaesthesia machine

Checking the integrity of the high pressure system

  1. Ensure the master switch is off, pipelines disconnected, cylinders completely closed and remaining gases in the system exhausted (pressure gauges reading zero). Confirm each cylinder by colour coding and label. Confirm proper attachment to the machine through the hanger yoke assembly.
  2. Open O 2 cylinder with cylinder key by full 3΍ rotations anticlockwise. Confirm the cylinder is at least half full (>7000 kPa or 1000 psi). Open O 2 flow control valve (anticlockwise rotation of knob) and confirm the O 2 flow meter registers flows (further confirmation of O 2 can be done by O 2 analyser). Open N 2 O flow control valve and confirm N 2 O flow meter doesn't register flows. Close the O 2 cylinder, wait until the flow reaches zero and the O 2 pressure gauge reads zero.
  3. Checking the integrity of N 2 O slave mechanism and oxygen pressure fail-safe mechanism: With the O 2 supply off, open N 2 O cylinder fully and confirm the N 2 O pressure gauge reads > 5000 kPa or 750 psi (lesser pressures mean exhaustion of liquid N 2 O). Open N 2 O flow control valve and confirm the absence of flow in the N 2 O flow meter (presence of flow indicates a defect). Open both O 2 cylinder and O 2 flow control valve (if not already open). Confirm both O 2 and N 2 O flow meters now register flows. Close the O 2 cylinder and flush the O 2 to confirm flows return to zero in both the O 2 and N 2 O flow meters.
  4. Oxygen flush should function even with the master switch and O 2 flow meter turned off as long as O 2 supply to the machine is ensured. The O 2 flush should stop as soon as the pressure on the O 2 flush knob is taken off.
  5. Close the O 2 and N 2 O cylinders and turn off the flow control valves for both gases.
Checking the integrity of the intermediate pressure system

  1. Tug test: Connect O 2 pipeline to the oxygen wall outlet using the Schrader quick coupler system. Correct coupling will not allow detachment of the pipeline from the Schrader coupler when a tug is given to the pipeline. Similar test can be performed with the N 2 O pipeline with N 2 O wall outlet.
  2. Single hose test: Disconnect N 2 O pipeline while retaining the O 2 pipeline intact. Open the O 2 flow control valve to note O 2 is flowing (further confirmation of O 2 can be done by oxygen analyser). Open N 2 O flow control valve which may show initial flows (residual N 2 O in the system) that subsequently falls to zero. Connect the N 2 O pipeline to its wall outlet and note again there is flow in the N 2 O flow meter. These steps help detect accidental mix up of O 2 and N 2 O pipeline connections. Disconnection of O 2 pipeline should result in both flow meters registering zero flows and activation of the oxygen fail-safe mechanism.
  3. Connect O 2 and N 2 O pipelines again and note their pressure gauges read >400 kPa or 55-60 psi (to ensure supply from the manifold room is at correct pressures).
Checking integrity of the low pressure system

This is performed after the following set up is established with the master switch on: Pipelines of O 2 and N 2 O intact, cylinders closed or in the absence of pipelines, cylinders open.

  1. Close the flow control valves, place vapourisers in their location on the machine with vapouriser dial turned off. Confirm sufficient liquid volatile agent and the filler cap is tightly shut. Ensure vapourisers are upright and not tilted (this prevents unsafe delivery of vapours).
  2. Universal negative pressure leak test: Turn master switch off and close all the flow control valves, attach suction bulb to the common gas outlet and repeatedly squeeze to empty its contents until the bulb is well collapsed. The bulb should remain collapsed for at least 10 s. To test for leaks in the vapourisers, individual vapouriser should be turned on and above mentioned steps be repeated. Re-inflation of the bulb within next 10 s indicates a leak in the low pressure system (when the vapouriser is off) or vapouriser (if an individual vapouriser is turned on during the test). At the end of this test, put the master switch on, remove the suction bulb and connect breathing apparatus.
  3. Open individual flow meters to their maximum range to confirm proper functioning of the Thorpe's tubes and the float. Confirm anti-hypoxic mechanisms are working satisfactorily through various ranges of O 2 and N 2 O flows.
Checking the electrical/electronic components of anaesthesia machine

  1. Turn the master switch on and confirm proper working of all other associated electrical or electronic equipment. If the machine has minimum mandatory flows, confirm O 2 flow meter registers a flow of around 50-200 mL with the O 2 flow control valve turned off.
  2. Confirm anaesthesia machine is connected to the mains (AC source) and the switch is on. Ensure the battery has at least 30 min back up supply and is charging during machine use.
  3. Monitor: Check appropriate functioning of SpO 2 , non-invasive blood pressure (NIBP), end tidal capnogram (ETCO 2 ), etc., by using them on self (e.g., SpO 2 on our finger >96%, exhalation into capnograph port registers a CO 2 waveform, etc.) and adjust alarm settings according to patient profile. Ensure monitoring equipment including gas sampling lines are secured leak free and kink free. Gas sampling line must be connected proximal to the airway filter to avoid frequent obstruction by moisture. Monitoring parameters not required for a given patient should be turned off.
  4. Oxygen analyser calibration (21% to >95%): Calibrate the analyser to read 21% at atmosphere. With the O 2 source from cylinder (pipeline source disconnected), open O 2 flow control valve and connect the analyser to the common gas outlet and calibrate to register at least 95%.
Checking other components of the anaesthesia machine

Most of the modern machines have the facility for connecting both circle system and Mapleson breathing system in such a way that just by turning a knob the desired breathing system can be put to use. Ensure the breathing circuit intended for use on the patient is correctly chosen (check the knob position).

  1. Circle system: Verify adequate fresh CO 2 absorbent and its proper attachment to the machine. Make all necessary connections of circle system components. Perform the leak test by occluding the patient end of the breathing circuit. Either increase O 2 flows or use O 2 flush to pressurise the breathing apparatus to >30 cm H 2 O. Turn the O 2 flow control valve off and stop O 2 flush. Drop in airway pressures to <30 cm H 2 O within 10 s indicate a leak in the system. Further quantification of the leak can be done by increasing the O 2 flows in small increments until the pressures can be sustained >30 cm H 2 O. The system pressure should be released by opening the APL valve. This ensures proper functioning of the APL valve and prevents accidental entry of the absorbent dust into the breathing system. Simultaneously, evaluate for appropriate response and functioning of the unidirectional valves.
  2. The integrity of individual Mapleson system should be tested but describing these is beyond the scope of this article.
  3. Ventilator: With the breathing system in situ and the patient end occluded, turn on the ventilator knob to evaluate the integrated ventilator. In case of ascending bellows, ensure the bellows reach the top of the bottle and then turn off the fresh gas flows. The bellows should continue to reach the top of the bottle at the end of each ventilator breath. Failure of the bellows to reach the top indicates leak. However, in case of descending bellows, this cannot be verified. Verify ventilator settings appropriate for the patient's weight and adjust alarm settings accordingly.
  4. Check appropriate connection of the scavenging system to the machine and its correct functioning.
  5. Ensure suction apparatus is working appropriately and sufficient negative pressures are rapidly developed when its port is occluded.
  6. Pay attention to any notes attached to workstation such as last servicing date, last time the CO 2 absorbent was changed, etc.
  7. When the breathing circuit is not in use, patient end must be covered with sterile layer. Common practice in our department is to place the patient end in a sterile glove.
Overall time taken for this protocol does not exceed 10 min. This duration might further be reduced if the machine is capable of self-check for its components. Between two anaesthetic conducts, any new equipment intended for use on next patient such as suction tubing, breathing circuitry, etc., should be tested. Verify sufficient availability of fresh CO 2 absorbent and volatile anaesthetic liquid. If O 2 cylinder was used for any reason, confirm the cylinder is at least half full or change to a new full cylinder. During the anaesthetic conduct, in the event of change of anaesthesiologist, proper hand over must be given regarding the machine check and functioning of all the components. In long duration procedures, periodically check for exhaustion of volatile anaesthetic liquid and CO 2 absorbent. A detailed check is required following any critical event if suspected to be due to workstation.

Clear documentation of regular and appropriate servicing of the anaesthesia machine, its components and their satisfactory functioning following servicing and repair is important. Finally, anaesthesiologist should be aware of manufacturing differences in the components and their functioning and should develop a machine check protocol convenient for their set up.


   Hazards of Anaesthesia Machines Top


Main problems related to older anaesthesia machines can broadly be classified into delivery of lower inspired oxygen concentrations, delivery of dangerously high or low concentrations of volatile anaesthetic agents, insufficient ventilation, excessive airway pressures, foreign bodies, hyperventilation and miscellaneous. [1],[6] Misfilling of cylinders, misconnections of pipelines or cylinders, delivery of 100% N 2 O (in the absence of hypoxic guard, N 2 O can be administered without opening oxygen flow control valve due to variations in the arrangement of O 2 flow meter in relation to others), leak at the flow meter assembly or vapourisers, etc., contributed to hypoxic gas delivery. [1],[6] To prevent this, several measures have been incorporated such as quick coupler system and diameter index safety system for pipelines, pin index safety system and colour coding for cylinders, O 2 failure safety devices (gas proportionating devices) to ensure minimum 25% O 2 supply by rendering the N 2 O a slave to O 2 pressures, O 2 gas downstream of other flow meters and use of O 2 analyser. The dangers related to vapourisers have been minimised by the incorporation of keyed index safety system for filling vapourisers and temperature-compensating technology while negative pressure leak test prior to use of vapourisers detects any leak in the vapourisers. Excessive pressure build-up was observed by obstruction in the pipeline, machine or circuitry to flow of gases by dust, blood, secretions, foreign bodies, stuck valves, etc., Insufficient ventilation could be attributed to leaks in the machine or breathing system. These have been minimised by incorporating integrated monitoring along with alarm systems for tidal volume, minute ventilation, airway pressures, etc., and diligent protocols for machine and breathing circuit check. [1],[6]

Modern machines have overcome many drawbacks associated with the older machines. However, addition of several mechanical, electronic and electric components has contributed to recurrence of some of the older problems such as leak or obstruction attributable to newer gadgets and development of newer problems. Our pubmed search for hazards associated with anaesthesia machines from the year 2000 onwards yielded several results indicating continued occurrence of machine related hazards. Although human errors cannot be attributed to machine, several problems were triggered by the modern machines and their components.

Hypoxic gas delivery due to problems with cylinders, pipelines, anaesthesia machines

Hypoxic gas delivery is still a distinct possibility. Development of a stricture in the O 2 central supply system outlet as a result of degradation of the O-ring and a structural defect in the pipeline delivery at the ceiling level of the operating room resulting in accidental switching off of the O 2 supply valve by the N 2 O pipeline have contributed to delivery of hypoxic gas mixtures. [7],[8] Misconnection of O 2 pipeline hose to N 2 O cylinder in the manifold room by technical personnel has resulted in hypoxic gas delivery. [9],[10] Insertion of the Equanox (50% each of O 2 and N 2 O) probe accidentally into the N 2 O wall outlet resulted in 100% N 2 O delivery. [11] These problems were compounded by either lack of oxygen analyser or failure to recognise the hypoxia early and changing over to an alternate plan by the concerned anaesthesiologist. Fault in the chain-link mechanism of Ohmeda Excel 210 SE where loosening of the stop screw which placed over the O 2 flow control knob contributed to hypoxic gas delivery. [12],[13],[14] Faulty interface between gear wheels of O 2 and N 2 O flow meters in ageing machines contributed to failure of the flow proportionating devices while defective rubber seal of flow meter control tube was responsible for hypoxic gas delivery. [15],[16]

Other reported hazards due to problems within the anaesthesia machine or workstation

Advances in technology contributed to the development of piston-driven ventilators in place of bellow operated ventilators. Resetting of a ventilator piston following suctioning of the airway during one-lung ventilation resulting in inability to ventilate is reported. [17] Improperly fitted retaining ring (placed between the expiratory valve assembly and the spiromed respiratory volume monitor) of the Narkomed 4 anaesthesia system contributed to gas leak. [18] Accidental obstruction to exhaust gas port of a ventilator by a vinyl bag resulted in ventilatory failure. [19] Awareness under anaesthesia occurred due to a disconnection of the common gas outflow tract prior to one way check valve of the anaesthesia machine while a to-and-fro type of anaesthesia ventilator operated without any problem. [20]

The APL valve has resulted in leak in the breathing system and kinking of the sampling ports. [21],[22],[23],[24],[25],[26],[27],[28],[29],[30] Improper seating of the vapourisers over the back bar, faulty locking spring on the vapouriser and broken transverse pin of the desflurane vapouriser contributed to gas leak. [31],[32],[33],[4],[35] During the process of filling of an isoflurane vapouriser, accidental spillage of the liquid caused corrosion and damage to the water trap while accidental lifting up of the lever of the Tec 5 isoflurane vapouriser resulted in damage to the water trap in Draeger machines. [36],[37],[38],[39],[40]

Hazards due to breathing systems and their use with newer anaesthesia machines

Hypercarbia and related problems are reported with faulty Bain's circuit and adult co-axial breathing circuits. [41],[42] Most of the modern machines have integrated circle system as one of their main components to enable the economy of gases and minimise atmospheric pollution. However, numerous possibilities for misconnections exist with circle system. In fact, a closed claims analysis revealed problems in breathing circuitry connections contributed to 35% adverse anaesthetic outcomes arising from gas delivery systems. [43] Problems in the CO 2 canister resulted in rebreathing and hypercarbia, [44],[45] significant leak and difficulty in ventilation, [46],[47],[48],[49],[50] and obstruction to ventilation. [51] Although use of circle system, CO 2 absorbents and low-flow anaesthesia are beneficial in economy of gases, they have inadvertently contributed to ventilator problem due to water condensation and production of dangerous substances such as compound A. [52],[53]

Inability to detect minor leaks in the machine or breathing circuit is observed with modern machines having minimum mandatory oxygen flow, while testing for leak with Bain's circuit was also found to be difficult with modern Aestiva 5 ® anaesthetic machine. [54],[55] Designing the ratio valve of the minimum mandatory O 2 flow system to vent to the atmosphere on switching the master switch off resulted in false positive for leaks when the universal leak test was applied in Cavendish anaesthesia machines. [56] Intraoperative replacement of reservoir bag by new latex-free bag contributed to difficulty in ventilation due to the presence of a large hole in the newly replaced bag. [57] Twisting of the bag around its own neck resulted in a tight bag scenario. [58] Increased depth of corrugations minimises circuit kinking; [59] however, this can result in leaks as damage to corrugated tubing is not easily identifiable on inspection. [60] Disposable breathing circuits and airway equipment minimise infective risks. However, transparent packaging of the breathing circuit accidentally caught in the straight connector of the breathing circuit was responsible for obstruction to ventilation. [61] Scavenging system reduces theatre pollution. However, scavenging equipment has contributed to obstruction to flow of gases. [62],[63],[64],[65]

Despite significant improvements in the design and safety aspects of anaesthesia machine and its components, some of the older and well-known problems continue to exist such as water vapour condensation inside the machine components, bobbins of the flow meters may get stuck to the inner wall of the flow meter due to dirt and static electricity, possibility of leak from the selectatec system of vapourisers in the event of accidental removal of O-ring during the process of mounting or dismounting of the vapourisers, etc.


   Scavenging Top


Long term exposure to trace anaesthetic gases released into the operating room during the conduct of general anaesthesia may be harmful to health-care personnel involved. There is evidence to show a higher rate of spontaneous abortion in women. Further, this could contribute to an increase in the incidence of infertility in the operating room personnel as well as higher chances of having children with congenital anomalies. Anaesthesiologists appear to have a predisposition to the development of various organ system disorders that may be attributable to long-term exposure to trace anaesthetic gases. [66] Therefore, the need for scavenging cannot be overemphasised.

Scavenging is the process of collection of exhaled gases from the anaesthetic equipment and disposal of the same to an appropriately designated place away from the operating room. Scavenging system is designed to minimise theatre pollution. Basic components necessary for efficient scavenging include a system to recover the exhaled gases from anaesthetic equipment, tubing to transfer these to a receiving reservoir and a system to dispose these gases away from the operating room. Ideal scavenging system components should be free from leaks and kinks, have a colour and diameter different from that of the conventional breathing circuitry to prevent misconnections. [66],[67]

Scavenging apparatus includes a system to recover exhaled gases from the anaesthetic equipment that should not cause resistance to exhalation and fit correctly over the exhaust port of the anaesthesia ventilator and the APL valve of the breathing circuits. Most receiving and reservoir systems in anaesthesia use are open types where the gases are collected into a reservoir and are transferred to the disposal system through a suction port. These also minimise transmission of pressure fluctuations in the scavenging system to anaesthetic circuitry. Open reservoir systems communicate to atmosphere through one or more ports. Closed systems communicate to atmosphere through valves and do not require a reservoir if active disposal system is used. The disposal apparatus can function either actively or passively. Patient's exhalation effort or positive pressure generated from manual ventilation or ventilator helps push the gases to their exit destination in the passive system. This can also be reached with the help of air circulating systems. Although activated charcoal canisters can be used to passively adsorb the volatile anaesthetic gases, they do not adsorb N 2 O and require frequent refilling. Catalytic decomposition of N 2 O to nitrogen and O 2 can be considered to minimise the harmful effects of N 2 O. In the active system, a working fan or vacuum pump draw the gases to their exit destination. Vacuum disposal creates the requirement for multiple vacuum ports (surgical, anaesthetic and scavenging) in the operating room and care to prevent negative pressures affecting the breathing system. Fans are less efficient than vacuum and require wide bore tubing. Open receiving and reservoir systems require active gas disposal system. [66],[67]

Other ways of minimising atmospheric pollution include gas delivery equipment with negligible leaks, low flow anaesthesia, minimal leak around the airway equipment (facemask, tracheal tube, laryngeal mask airway, etc.), ≥15 air changes per hour and total intravenous anaesthesia. [66],[67]


   Summary Top


The review highlights the fact that problems can occur despite the incorporation of several safety aspects to anaesthesia machine. Human factors have contributed to greater complications than machine faults. Therefore, better understanding of the basics of anaesthesia machine and checking each component of the machine for proper functioning prior to use is essential to minimise these hazards. Despite advanced technology, a remote but life-threatening possibility of intraoperative machine malfunction exists. A self-inflating bag appropriate for the patient's age and alternate O 2 source should be present as rescue measures in the event of machine malfunction.

 
   References Top

1.Hazards of anesthesia machines and breathing systems. In: Dorsch JA, Dorsch SE, editors. Understanding Anesthesia Equipment. 5 th ed. Philadelphia, USA: Lippincott Williams and Wilkins; 2007. p. 373-403.  Back to cited text no. 1
    
2.Merchant R, Chartrand D, Dain S, Dobson G, Kurrek M, Lagacé A, et al. Guidelines to the practice of anesthesia revised edition 2013. Can J Anaesth 2013;60:60-84.  Back to cited text no. 2
    
3.Hartle A, Anderson E, Bythell V, Gemmell L, Jones H, McIvor D. et al. Checking anaesthetic equipment 2012: Association of anaesthetists of great Britain and Ireland. Anaesthesia 2012;67:660-8.  Back to cited text no. 3
    
4.Australian and New Zealand College of Anaesthetists. Recommendations on checking anaesthesia delivery systems, 4.2.3.4.2, 2003. Available from: http://www.anzca.edu.au/resources/professional-documents/ps31.html. [Last accessed on 2013 May 30].  Back to cited text no. 4
    
5.Recommendations for Pre-Anesthesia Checkout Procedures. Sub-committee of ASA committee on equipment and facilities, 2008. Available from: http:// www.asahq.org/For-Members/Clinical-Information/2008-ASA- Recommendations-for-PreAnesthesia- Checkout.aspx. [Last accessed on 2013 May 30].  Back to cited text no. 5
    
6.Brockwell RC, Andrews JJ. Complications of inhaled anesthesia delivery systems. Anesthesiol Clin North America 2002;20:539-54.  Back to cited text no. 6
    
7.Desaki Y, Yorozuya T, Nakanishi K, Soutani M, Nagaro T. Stricture of oxygen outlet of the central piping identified by a decrease in the oxygen supply pressure into the anesthesia machine. Masui 2011;60:507-10.  Back to cited text no. 7
    
8.Sugiuchi N, Miyazato K, Hara K, Horiguchi T, Shinozaki K, Aoki T. Failure of operating room oxygen delivery due to a structural defect in the ceiling column. Masui 2000;49:1165-8.  Back to cited text no. 8
    
9.Herff H, Paal P, von Goedecke A, Lindner KH, Keller C, Wenzel V. Fatal errors in nitrous oxide delivery. Anaesthesia 2007;62:1202-6.  Back to cited text no. 9
    
10.Herff H, Paal P, Lindner KH, von Goedecke A, Keller C, Wenzel V. Fatalities due to nitrous oxide. Complications from mistakes in nitrous oxide supply. Anaesthesist 2008;57:1006-10.  Back to cited text no. 10
    
11.McKenzie AG, McClure JH, Scott DH, Amoore JN, Leitch S. Insertion of equanox probe into terminal outlet for nitrous oxide with consequent hypoxia. Anaesthesia 2009;64:1375-6.  Back to cited text no. 11
    
12.Cheng CJ, Garewal DS. A failure of the chain-link mechanism on the Ohmeda Excel 210 anesthetic machine. Anesth Analg 2001;92:913-4.  Back to cited text no. 12
    
13.Paine GF, Kochan JJ 3 rd . Failure of the chain-link mechanism of the Ohmeda Excel 210 anesthesia machine. Anesth Analg 2002;94:1374.  Back to cited text no. 13
    
14.Sharma ML. Problem with Ohmeda Excel 210 SE anesthetic machine. Can J Anaesth 2002;49:438-9.  Back to cited text no. 14
    
15.Ishikawa S, Nakazawa K, Makita K. Hypoxic gas flow caused by malfunction of the proportioning system of anesthesia machines. Anesth Analg 2002;94:1672.  Back to cited text no. 15
    
16.Hay H. Delivery of an hypoxic gas mixture due to a defective rubber seal of a flowmeter control tube. Eur J Anaesthesiol 2000;17:456-8.  Back to cited text no. 16
    
17.Barahal D, Sims C. Ventilator failure during use of a new anesthesia machine. Anesthesiology 2001;95:568-9.  Back to cited text no. 17
    
18.Chung DC, Ho AM, Tay BA. "Apnea-volume" warning during normal ventilation of the lungs: An unusual leak in the Narkomed 4 Anesthesia System. J Clin Anesth 2001;13:40-3.  Back to cited text no. 18
    
19.Ushijima K, Yamada Y, Sasao Y, Ichinose K. Anesthesia machine malfunction caused by the blockade of an exhaust gas port. Masui 2007;56:200-2.  Back to cited text no. 19
    
20.Inoue S, Sakamoto T, Kitaguch K, Furuya H. Awareness due to disconnection from the fresh gas supply: Why could ventilation be achieved in spite of disconnection from the fresh gas supply? Acta Anaesthesiol Sin 2000;38:85-8.  Back to cited text no. 20
    
21.Souvatzis X, Askitopoulou H. Malfunction of an adjustable pressure limit valve. Eur J Anaesthesiol 2007;24:978-80.  Back to cited text no. 21
    
22.Hennenfent S, Suslowicz B. Circuit leak from capnograph sampling line lodged under adjustable pressure limiting valve. Anesth Analg 2010;111:578.  Back to cited text no. 22
    
23.Robards C, Corda D. A potential hazard involving the gas sampling line and the adjustable pressure limiting valve on the Drager Apollo anesthesia workstation. Anesth Analg 2010;111:578-9.  Back to cited text no. 23
    
24.Kibelbek MJ. Cable trapped under Dräger Fabius automatic pressure limiting valve causes inability to ventilate. Anesthesiology 2007;106:639-40.  Back to cited text no. 24
    
25.Vijayakumar A, Saxena DK, Sivan Pillay A, Darsow R. Massive leak during manual ventilation: Adjustable pressure limiting valve malfunction not detected by pre-anesthetic checkout. Anesth Analg 2010;111:579-80.  Back to cited text no. 25
    
26.Oprea AD, Ehrenwerth J, Barash PG. A case of adjustable pressure-limiting (APL) valve failure. J Clin Anesth 2011;23:58-60.  Back to cited text no. 26
    
27.Sanders R. A new APL valve hazard. Anaesthesia 2001;56:1119-20.  Back to cited text no. 27
    
28.Hirabayashi G, Uchino H, Ogihara Y, Omi A, Ohmura A. Inadequate gas supply to patients with an adjustable pressure-limiting valve in the fully opened position. J Anesth 2010;24:626-9.  Back to cited text no. 28
    
29.Eggemann I, Böttiger BW, Spöhr F. Accidentally opened adjustable pressure-limiting valve. Failure of manual ventilation. Anaesthesist 2009;58:301-2.  Back to cited text no. 29
    
30.Kojima O, Kuzukawa A, Sai Y. Inadequate gas supply to the patients after the repair of anesthesia delivery system. Masui 2010;59:1405-7.  Back to cited text no. 30
    
31.Wilkinson D, Curtis C. A strange leak from the anaesthetic machine. Anaesthesia 2002;57:1038-9.  Back to cited text no. 31
    
32.Ong BC, bin Katjo J, Tan BL, Lee CC, Chan YW. Acute failure of oxygen delivery. Anesthesiology 2001;95:1038-9.  Back to cited text no. 32
    
33.Garstang JS. Gas leak from TEC 5 isoflurane vaporiser. Anaesthesia 2000;55:915.  Back to cited text no. 33
    
34.Terry L, da Silva EJ. Faulty spring causing a gas leak. J Clin Anesth 2009;21:382-3.  Back to cited text no. 34
    
35.Liew WL, Jayamaha J. Anaesthetic machine leak from desflurane vaporiser. Anaesthesia 2011;66:399-400.  Back to cited text no. 35
    
36.Vemmer T, Srikiran R, Liban B. Isoflurane damage to a Draeger primus anaesthetic machine. Anaesthesia 2004;59:306-7.  Back to cited text no. 36
    
37.McAllister RK, Payne MN, Bittenbinder TM. Disruption of water trap from leak of isoflurane during the vaporizer filling process in the Dräger Apollo. Anesthesiology 2007;107:514.  Back to cited text no. 37
    
38.Reddy S, Bishop SM, Moore JA. Ohmeda Tec 5 isoflurane vaporiser and damage to the water trap on the Draeger primus anaesthesia machine. Anaesthesia 2008;63:892-3.  Back to cited text no. 38
    
39.Rajendram R, Laurie P, Willatts D. Potential danger hidden behind vaporizers on Datex-Ohmeda excel 210 SE. Eur J Anaesthesiol 2007;24:1068-9.  Back to cited text no. 39
    
40.Kimatian SJ. With technology comes responsibility: Intraoperative failure of an anesthetic vaporizer. Anesthesiology 2002;96:1533-4;1534.  Back to cited text no. 40
    
41.Ghai B, Makkar JK, Bhatia A. Hypercarbia and arrhythmias resulting from faulty Bain circuit: A report of two cases. Anesth Analg 2006;102:1903-4.  Back to cited text no. 41
    
42.Jellish WS, Nolan T, Kleinman B. Hypercapnia related to a faulty adult co-axial breathing circuit. Anesth Analg 2001;93:973-4.  Back to cited text no. 42
    
43.Caplan RA, Vistica MF, Posner KL, Cheney FW. Adverse anesthetic outcomes arising from gas delivery equipment: A closed claims analysis. Anesthesiology 1997;87:741-8.  Back to cited text no. 43
    
44.Agrawal P, Gupta B, D'souza N. An unusual cause of carbon dioxide rebreathing in a circle absorber system. J Anesth 2010;24:976-7.  Back to cited text no. 44
    
45.Lin SL, Yeh CC, Lu CH, Ho ST, Wong CS. Hypercapnia due to rupture of the unidirectional valve in the inspiratory limb of the breathing system after induction of general anesthesia - A case report. Acta Anaesthesiol Sin 2003;41:139-43.  Back to cited text no. 45
    
46.Umesh G, Jasvinder K, Sagarnil R. Leak in the breathing circuit: CO2 absorber and human error. J Clin Monit Comput 2010;24:143-4.  Back to cited text no. 46
    
47.Wax D, Neustein S. Malfunction of the new Aisys anesthesia machine. Anesthesiology 2007;106:404-5.  Back to cited text no. 47
    
48.Sumi C, Asai T, Kawashima A, Nawa T, Shingu K. Gas leakage from an anesthetic circuit caused by mis-installation of a canister. Masui 2008;57:1427-30.  Back to cited text no. 48
    
49.Bader SO, Doshi KK, Grunwald Z. A novel leak from an unfamiliar component. Anesth Analg 2006;102:975-6.  Back to cited text no. 49
    
50.Paul C, Böttiger BW. Cleft in carbon dioxide absorber. Intraoperative problems with ventilation due to a leak in the breathing circuit. Anaesthesist 2010;59:652-4.  Back to cited text no. 50
    
51.Bagdasarian LA, Shah NK, Sharif MK. Carbon dioxide absorber as a cause of high airway pressures. J Clin Anesth 2008;20:48-9.  Back to cited text no. 51
    
52.Kim KO. Ventilator malfunction due to water condensation during low flow anaesthesia. Anaesth Intensive Care 2011;39:1155-6.  Back to cited text no. 52
    
53.Yamakage M, Kimura A, Chen X, Tsujiguchi N, Kamada Y, Namiki A. Production of compound A under low-flow anesthesia is affected by type of anesthetic machine. Can J Anaesth 2001;48:435-8.  Back to cited text no. 53
    
54.Tokumine J, Sugahara K, Gushiken K, Ohta M, Matsuyama T, Saikawa S. Non-zero basal oxygen flow a hazard to anesthesia breathing circuit leak test. Anesth Analg 2005;100:1056-8.  Back to cited text no. 54
    
55.Ip JK, Fauvel NJ. Difficulty with leak-testing Bain breathing systems on an Aestiva 5 anaesthetic machine. Anaesthesia 2008;63:321-2.  Back to cited text no. 55
    
56.Bhagwat AG, Saxena K, Narang S, Jain S. 'Universal' leak test? Anaesth Intensive Care 2008;36:912-3.  Back to cited text no. 56
    
57.Sandberg WS, Kaiser S. Novel breathing circuit architecture: New consequences of old problems. Anesthesiology 2004;100:755-6.  Back to cited text no. 57
    
58.Umesh G, Jasvinder K. Tight reservoir bag: The bag itself may be the culprit. J Clin Monit Comput 2010;24:189-90.  Back to cited text no. 58
    
59.Umesh G, Jasvinder K, Nanda S. Solutions to some of the existing problems with breathing equipment type. Acta Anaesthesiol Taiwan 2010;48:91-3.  Back to cited text no. 59
    
60.Rossberg MI, Greenberg RS. Anesthesia respiratory circuit failure. Anesthesiology 2002;97:762-3.  Back to cited text no. 60
    
61.Chacon AC, Kuczkowski KM, Sanchez RA. Unusual case of breathing circuit obstruction: Plastic packaging revisited. Anesthesiology 2004;100:753.  Back to cited text no. 61
    
62.Joyal JJ, Vannucci A, Kangrga I. High end-expiratory airway pressures caused by internal obstruction of the Draeger Apollo ® scavenger system that is not detected by the workstation self-test and visual inspection. Anesthesiology 2012;116:1162-4.  Back to cited text no. 62
    
63.Gray JS, Evans GA. Scavenging malfunction. Anaesthesia 2003;58:1144-5.  Back to cited text no. 63
    
64.Elakkumanan LB, Vasudevan A, Krishnappa S, Pandey RR, Balachander H, Badhe AS. Obstruction to scavenging system tubing. J Anaesthesiol Clin Pharmacol 2012;28:270-1.  Back to cited text no. 64
[PUBMED]  Medknow Journal  
65.Hay H. Oxygen delivery failure resulting from interference with a Bain breathing system. Eur J Anaesthesiol 2000;17:591-3.  Back to cited text no. 65
    
66.Controlling trace gas levels. In: Dorsch JA, Dorsch SE, editors. Understanding Anesthesia Equipment. 5 th ed. Philadelphia, USA: Lippincott Williams and Wilkins; 2007. p. 404-30.  Back to cited text no. 66
    
67.Mackenzie M. Anaesthetic gas scavenging. Available from: http://www.update.anaesthesiologists.org/2008/12/01/anaesthetic-gas-scavenging/. [Last accessed on 2013 May 31].  Back to cited text no. 67
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Anaesthesia Mach...
    Hazards of Anaes...
   Scavenging
   Summary
    References
    Article Figures

 Article Access Statistics
    Viewed6548    
    Printed59    
    Emailed4    
    PDF Downloaded2707    
    Comments [Add]    

Recommend this journal