Indian Journal of Anaesthesia  
About us | Editorial board | Search | Ahead of print | Current Issue | Past Issues | Instructions
Home | Login  | Users Online: 2116  Print this pageEmail this pageSmall font sizeDefault font sizeIncrease font size    




 
 Table of Contents    
LETTER TO EDITOR
Year : 2014  |  Volume : 58  |  Issue : 4  |  Page : 495-496  

Common source but a rare cause for intraoperative breathing circuit leak: Every anaesthesiologist should be aware of


Department of Anaesthesiology, PSG Institute of Medical Sciences and Research, Coimbatore, Tamil Nadu, India

Date of Web Publication17-Aug-2014

Correspondence Address:
Dr. Vinodhadevi Vijayakumar
Department of Anaesthesiology, PSG Institute of Medical Sciences and Research, Coimbatore - 641 004, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.139027

Rights and Permissions

How to cite this article:
Govindarajulu D, Vijayakumar V, Velusamy P. Common source but a rare cause for intraoperative breathing circuit leak: Every anaesthesiologist should be aware of. Indian J Anaesth 2014;58:495-6

How to cite this URL:
Govindarajulu D, Vijayakumar V, Velusamy P. Common source but a rare cause for intraoperative breathing circuit leak: Every anaesthesiologist should be aware of. Indian J Anaesth [serial online] 2014 [cited 2019 Oct 20];58:495-6. Available from: http://www.ijaweb.org/text.asp?2014/58/4/495/139027

Sir,

A 60-year-old American Society of Anaesthesiologists physical status II patient was posted for laparoscopic cholecystectomy. After protocolised anaesthesia machine check and preoxygenation using closed circuit, patient was induced and intubated. Correct placement of endotracheal tube (ETT) was confirmed with square wave capnography in bag mode. Circuit was switched to ventilator mode with fresh gas flows (FGF) of 2 L/min. Good square wave capnography was seen. A number 16 French nasogastric (NG) tube insertion was passed freely with minimal resistance. The NG tube insertion produced gurgling noise and on suctioning it was negative for gastric contents. Subsequently, we noticed inadequate filling of ventilator bellows.

Immediately, we increased the FGF to 6 L/min and simultaneously searched for the source of leak from all parts of the circle system. The capnography trace was unsatisfactory and expiratory tidal volume was displayed as 25% of set tidal volume. The circuit was changed to manual mode, but the leak persisted. Cause for the circuit leak at patient end was ruled out by reconfirming the position and cuff pressure of ETT. The patient was shifted to Bain breathing system, with FGF of 8 L/min. Oxygenation and capnography were satisfactory and hence, we proceeded with the surgery. Since the machine and circuit were assessed to be leak proof, we were suspicious about the NG tube position. We inserted the open end of NG tube into a water seal and noticed the gas bubbles escaping cyclically with respiration. When the open end of the NG tube was closed, the circle system was delivering the set tidal volume with normal square wave capnography. Thus we confirmed the source of circuit leak to be the NG tube malposition. Direct laryngoscopy confirmed the passage of the NG tube through the glottis, and it was removed. Extubation and postoperative period was uneventful.

Nasogastric tube is normally inserted after induction and intubation in elective surgeries. In the present case, after the NG tube insertion, the 'circuit leak' was recognised by the workstation sensors within few breaths, which diverted our focus to ensure adequate oxygenation and ventilation, rather than checking the NG tube position. The higher FGF used in Bain breathing system masked the circuit leak. NG tube malposition is associated with significant morbidity and mortality in intensive care unit. [1] The reported incidence of NG tube malposition into the tracheobronchial tree is reported to be 2%. [2]

Most of the complications due to malpositioning are related to feeding which is unlikely in the intra-operative period. Hence, the reported incidence of NG tube malposition in the intraoperative period are limited. [3] Direct laryngoscopy can help in identifying the relationship of the NG tube with the ET tube; radiological confirmation of the NG tube tip within the stomach shadow is confirmatory, which is not always feasible during the intra-operative period. Fuchs et al. mention the increase in cuff pressure as one of the non-radiological methods of confirming the malposition of NG tube into the trachea. [4]

The check for circuit leak should not end with the assessment of ET tube (correct tip position, cuff position and pressure). It should be extended to rule out NG tube malposition, especially when the circuit leak occurs immediately after its insertion. Inserting NG tube under direct vision using laryngoscopy could have prevented this complication. Routine protocolised anaesthesia machine check-up, back up ventilator equipments, prompt recognition of the problem and seeking help earlier would be safe.


   Acknowledgments Top


Anaesthesia technicians, Department of Anaesthesiology, PSG Institute of Medical Sciences and Research, Coimbatore.

 
   References Top

1.Tenholder MF, Erwin WA, Nelson HS. "Lost" tidal volumes in a 71-year-old ventilated patient. Chest 1994;106:1869-71.  Back to cited text no. 1
    
2.Rassias AJ, Ball PA, Corwin HL. A prospective study of tracheopulmonary complications associated with the placement of narrow-bore enteral feeding tubes. Crit Care 1998;2:25-28.  Back to cited text no. 2
    
3.Soni KD, Gupta B, Agrawal P, D'souza N, Sinha C. An uncommon cause of intraoperative airleak. Indian J Crit Care Med 2011;15:237-8.  Back to cited text no. 3
[PUBMED]  Medknow Journal  
4.Fuchs J, Schummer C, Giesser J, Bayer O, Schummer W. Detection of tracheal malpositioning of nasogastric tubes using endotracheal cuff pressure measurement. Acta Anaesthesiol Scand 2007;51:1245-9.  Back to cited text no. 4
    



This article has been cited by
1 Intraoperative Gastric Tube Intubation: A Summary of Case Studies and Review of the Literature
Michael Long,Melissa Machan,Luis Tollinche
Open Journal of Anesthesiology. 2017; 07(03): 43
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   Acknowledgments
    References

 Article Access Statistics
    Viewed922    
    Printed10    
    Emailed0    
    PDF Downloaded318    
    Comments [Add]    
    Cited by others 1    

Recommend this journal