LETTER TO EDITOR
Year : 2014  |  Volume : 58  |  Issue : 4  |  Page : 497-499  

Airway obstruction during one lung ventilation: A shocking twist in the tube


Department of Anaesthesiology and Critical Care, Command Hospital (EC), Kolkata - 700 027, West Bengal, India

Date of Web Publication17-Aug-2014

Correspondence Address:
S Kiran
Department of Anaesthesiology and Critical Care, Command Hospital (EC), Kolkata - 700 027, West Bengal
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.139029

Rights and Permissions

How to cite this article:
Ahluwalia C S, Kiran S, Chopra V, Kar S. Airway obstruction during one lung ventilation: A shocking twist in the tube. Indian J Anaesth 2014;58:497-9

How to cite this URL:
Ahluwalia C S, Kiran S, Chopra V, Kar S. Airway obstruction during one lung ventilation: A shocking twist in the tube. Indian J Anaesth [serial online] 2014 [cited 2019 Oct 18];58:497-9. Available from: http://www.ijaweb.org/text.asp?2014/58/4/497/139029

Sir,

Intraoperative airway obstruction is not an uncommon life threatening emergency. We encountered an incident of unexpected intraoral kinking of double lumen tube (DLT) intraoperatively during one lung ventilation (OLV) in right lateral decubitus position.

A 39-year-old male, diagnosed case of Pott's spine of T12-L1 vertebrae with paraparesis was posted for left sided thoracotomy and corpectomy with fusion of T12-L1 vertebrae. Patient was planned for OLV with DLT for better surgical exposure and was intubated with 39 Fr left sided DLT (SHER-I-BRONCH ® ; HUDSON RCI, CA, USA) which was fixed at 28 cm mark at the incisors. The patient was turned to right lateral decubitus position and optimal placement of DLT was rechecked with auscultation and fiberoptic bronchoscopy. After 1΍ h of surgery, left lung was collapsed, OLV of dependent right lung was instituted with a tidal volume of 500 ml. After 30 min of OLV, peak airway pressure increased from 24 to 32 cm H 2 O and expired tidal volume decreased to 400 ml. Peak airway pressure further increased to 45 cm H 2 O and expired tidal volume decreased to 100 ml over the next few minutes. End-tidal carbon dioxide rose to 78 mmHg, oxygen saturation dropped to 44%, hypertension and tachycardia were noted. Auscultation revealed absent breath sounds on the right side of the chest. Pressure controlled ventilation was instituted but expired tidal volumes were 100 ml at 40 cm H 2 O pressure. Patient was then ventilated manually with reservoir bag, oxygen saturation increased to 92% but there was a high resistance to bag squeeze. A tube block due to secretions or displacement was considered and suction catheter was passed into the tracheal lumen for suctioning. However, the suction catheter could not be negotiated more than 10 cm. The bronchial lumen was attached to the ventilator circuit and dual lung ventilation was tried. However, the collapsed left lung failed to expand inspite of manual ventilation with high inspiratory tidal volumes. Arterial blood gas analysis revealed severe respiratory acidosis with pH of 6.94, pCO 2 of 121 mmHg, bicarbonate of 26 mmol/L and pO 2 of 148 mmHg. Fiberoptic bronchoscopy was contemplated but was not feasible in this life threatening scenario.

The decision to convert to single lumen endotracheal tube was taken quickly, patient was turned to the supine position, DLT was removed and trachea was reintubated with a 8.5 mm internal diameter cuffed polyvinyl chloride (PVC) endotracheal tube. Adequate ventilation was achieved with air entry present on auscultation over the right lung and left lung expanded visibly. Expired tidal volume, airway pressure and capnometry normalised quickly and repeat arterial blood gas sample showed a pH of 7.44 with a pCO 2 of 46 mmHg. Surgery was continued with retraction of the left lung and completed. Neuromuscular blockade was reversed, patient was extubated and had an uneventful recovery.

Inspection of the DLT after extubation showed a sharp kink along the concavity at around 18 cm mark which was occluding both the lumens. The kink was in the oropharynx at the point where the patients' oral axis and the pharyngeal axis met.

Kinking of the endotracheal tube intraoperatively, though uncommon creates a life-threatening emergency. [1] Kinking generally occurs in PVC tubes during head and neck surgeries due to excessive bending of the neck and are usually extraoral. However, kinking can occur in reinforced tubes and DLTs and can be extraoral, intraoral or intratracheal. Endotracheal kinking of a DLT may occur as a potential complication of inappropriate DLT size selection. [2]

Apart from kinking of the DLT, intraoperative difficulty in ventilation during DLT use may also result from obstruction of the DLT by blood, secretions, foreign body, abnormal anatomy, endobronchial mass and external bronchial compression. Poor pulmonary or chest wall compliance, acute bronchospasm, tension pneumothorax also manifest as difficulty in ventilation. [3] Kinking and other causes of obstruction may present as an inspiratory wheeze, but bronchospasm presents as an expiratory wheeze. However, the most common cause of airway obstruction during DLT use is malposition. It occurs in about one-third of patients while positioning in lateral decubitus position; fiberoptic bronchoscopy to confirm tube placement after positioning is essential. [4]

The sudden unforeseen kinking after 2 h in this case at the junction of the oral and pharyngeal axis was caused by thermal softening of the PVC DLT due to the temperature inside the oral cavity. The tube softens and kinking occurs inside the oropharynx due to over bending [Figure 1]. Bending of the PVC DLT along the concave curve at room temperature does not compromise the tube patency but at body temperature kinking and lumen obstruction can occur at a low angle due to tube softening. [5] Medical grade silicon DLTs (Silbronco® DLT; Fuji Systems, Tokyo, Japan) have properties of less tube softening with prolonged use and have a reinforced endobronchial tip to reduce kinking; however further studies are necessary. [6] Failure to secure free airway following airway obstruction due to DLT kinking may be lethal. Early detection and a low threshold for extubation and initiation of rescue ventilation with another DLT, endotracheal tube or laryngeal mask airway is vital.
Figure 1: Double lumen tube showing a sharp intraoral kink due to thermal softening

Click here to view


 
   References Top

1.Szekely SM, Webb RK, Williamson JA, Russell WJ. The Australian Incident Monitoring Study. Problems related to the endotracheal tube: An analysis of 2000 incident reports. Anaesth Intensive Care 1993;21:611-6.  Back to cited text no. 1
    
2.Ambrosio C, Leykin Y, Pellis T, Brodsky JB. Endotracheal kinking of a double-lumen tube: A potential complication of inappropriate size tube selection. Eur J Anaesthesiol 2011;28:607-8.  Back to cited text no. 2
    
3.Choi DK, Hwang JH, Song MH, Hahm KD. Kinking of a left-sided double-lumen tube within the trachea. J Cardiothorac Vasc Anesth 2011;25:1119-20.  Back to cited text no. 3
    
4.Hilal S, Ugur G, Ali A, Cilsem SD, Ozlem G, Eser S. Comparison of two different right-sided double-lumen tubes with different designs. Turk J Med Sci 2012;42:1063-9.  Back to cited text no. 4
    
5.Hübler M, Petrasch F. Intraoperative kinking of polyvinyl endotracheal tubes. Anesth Analg 2006;103:1601-2.  Back to cited text no. 5
    
6.Lohser J, Brodsky JB. Silbronco double-lumen tube. J Cardiothorac Vasc Anesth 2006;20:129-31.  Back to cited text no. 6
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed1025    
    Printed10    
    Emailed0    
    PDF Downloaded272    
    Comments [Add]    

Recommend this journal