LETTER TO EDITOR
Year : 2016  |  Volume : 60  |  Issue : 12  |  Page : 963-964  

Difficulty in diagnosing physical damage to the airway tube of the ProSeal laryngeal mask airway


Department of Anaesthesia, Dharwad Institute of Mental Health and Neurosciences, Dharwad, Karnataka, India

Date of Web Publication9-Dec-2016

Correspondence Address:
Goneppanavar Umesh
Department of Anaesthesia, Dharwad Institute of Mental Health and Neurosciences, Dharwad, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.195504

Rights and Permissions

How to cite this article:
Jasvinder K, Umesh G. Difficulty in diagnosing physical damage to the airway tube of the ProSeal laryngeal mask airway. Indian J Anaesth 2016;60:963-4

How to cite this URL:
Jasvinder K, Umesh G. Difficulty in diagnosing physical damage to the airway tube of the ProSeal laryngeal mask airway. Indian J Anaesth [serial online] 2016 [cited 2017 Jan 16];60:963-4. Available from: http://www.ijaweb.org/text.asp?2016/60/12/963/195504

Sir,

A 27-month-old boy weighing 12 kg was scheduled for cystoscopy and evaluation. Preoperative evaluation was unremarkable. In the operating room, standard anaesthesia monitoring was established. Following intravenous (IV) induction of anaesthesia, a size 2 ProSeal laryngeal mask airway (ProSeal LMA™; The Laryngeal Mask Company Limited, Osprey House, Old Street, St. Helier, Jersey JE2 3RG Channel Islands) was inserted and the position confirmed by observing coordinated movement of the reservoir bag with spontaneous breathing efforts of the child. Anaesthesia was maintained with sevoflurane in a mixture of oxygen and nitrous oxide with the child breathing spontaneously. Subsequently, the child was shifted to the distal end of the operating table and placed in lithotomy position. Till this time, the child was breathing spontaneously, and a good capnogram trace (EtCO 2 ) was observed. When the cystoscope was inserted, the child started making grunting noises. A propofol bolus (20 mg IV) was administered following which the child developed apnoea. Manual positive pressure ventilation (PPV) with both the circle system and Jackson Rees modification of Ayre's T-piece revealed the absence of observable chest expansion, absence of EtCO 2 and inability to fill the reservoir bag despite occluding the expiratory valve completely. We suspected a significant leak or a displaced laryngeal mask airway (LMA). Further evaluation of the breathing circuit and the LMA revealed a horizontal slit in the airway tube of the LMA [Figure 1]a. In view of this, we considered either replacing the LMA with another one or performing tracheal intubation. However, since the child was in an unfavourable position for airway instrumentation, we passed a well lubricated uncuffed 4.0 mm ID tracheal tube into the airway tube of the ProSeal LMA™ till its tip bypassed the portion that contributed to leak. The breathing circuit was then connected to the tracheal tube which enabled us to provide PPV through the ProSeal LMA™. One anaesthesia technician held the tracheal tube and the LMA together till the surgical evaluation was completed. Further management was uneventful.
Figure 1: (a) A normal appearing size 2 ProSeal LMA™ on external evaluation (left image) and the same ProSeal LMA™ when bent reveals significant crack just below its bite block portion (right image) (b) A normal appearing size 21/2 ProSeal LMA™ on external evaluation (left image) and the same ProSeal LMA™ when bent reveals two tiny cracks just below its bite block portion (right image). LMA - Laryngeal mask airway

Click here to view


Physical damage to reinforced tubes resulting in intraoperative leak and inability to ventilate is a well-recognised complication necessitating their replacement with either another supraglottic device or by a tracheal tube. [1],[2],[3] However, we managed our situation in a unique way because of the possibility of difficulty in handling the airway as the child was placed midway on the operating table in the lithotomy position. Prior knowledge about the usefulness of a ProSeal LMA™ for intubation and correct knowledge of the maximum size tracheal tube that might pass through a size 2 ProSeal LMA™ helped us in rapidly resolving the problem.

As per the current practice to ensure safety in cases of reinforced tubes, [1],[2],[3],[4] internal as well as external evaluation of the reinforced airway lumen of ProSeal LMA™ was done by both the anaesthesia technician and the anaesthesiologist preoperatively. However, both could not identify the damage because the horizontal slit in the lumen of the reinforced tube was difficult to notice when the device was in its anatomical position. As a result of this experience, we started gently flexing the reinforced portion of such devices in multiple directions during preoperative check. In the following 3 months, another ProSeal LMA™ [Figure 1]b with damage at the same location as the previous one (just beneath the bite block portion of the airway tube) was identified. When soap solution was applied at the damage site with 6 L oxygen flow passing through the ProSeal LMA™ with its laryngeal end partially occluded, bubbling of gases through the damaged portion confirmed leak. The damage could be because the concerned devices on both occasions were used >50 times.

We suggest that the lumen of reinforced tubes be bent in multiple directions preoperatively to enhance the possibility of detecting any physical damage to the device.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Matioc A, Arndt GA. Intubation using the ProSeal LMA™ laryngeal mask airway and a cook airway exchange catheter set. Can J Anaesth 2001;48:932.  Back to cited text no. 1
    
2.
Ladi SD, Aphale S. Accidental transection of flexometallic endotracheal tube during partial maxillectomy. Indian J Anaesth 2011;55:284-6.  Back to cited text no. 2
[PUBMED]  Medknow Journal  
3.
McLean R, Houston P, Carmichael F, Bernstein M. Disruption of an armoured endotracheal tube caused by biting. Can Anaesth Soc J 1985;32 (3 Pt 1):313-4.  Back to cited text no. 3
    
4.
Mesa A, Miguel R. Hidden damage to a reinforced LMA-Fastrach™ endotracheal tube. Anesth Analg 2000;90:1250-1.  Back to cited text no. 4
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed281    
    Printed1    
    Emailed0    
    PDF Downloaded55    
    Comments [Add]    

Recommend this journal