BRIEF COMMUNICATION
Year : 2017  |  Volume : 61  |  Issue : 6  |  Page : 511-512  

Airway ultrasound to detect endotracheal tube cuff herniation


Department of Anaesthesiology and Critical Care, Indira Gandhi Medical College and Research Institute, Puducherry, India

Date of Web Publication12-Jun-2017

Correspondence Address:
Sangeeta Dhanger
FR4, Sri Anbalaya Apartments, 17th Cross Street, Krishna Nagar, Puducherry - 605 008
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ija.IJA_15_17

Rights and Permissions

How to cite this article:
Tripathy DK, Bhat R R, Dhanger S. Airway ultrasound to detect endotracheal tube cuff herniation. Indian J Anaesth 2017;61:511-2

How to cite this URL:
Tripathy DK, Bhat R R, Dhanger S. Airway ultrasound to detect endotracheal tube cuff herniation. Indian J Anaesth [serial online] 2017 [cited 2019 Dec 10];61:511-2. Available from: http://www.ijaweb.org/text.asp?2017/61/6/511/207756




   Introduction Top


Endotracheal tube (ETT) cuff herniation is not an uncommon event in Intensive Care Units and operating rooms. The cause for this can be displacement of the tube or overinflation of ETT cuff.[1] Usually, it is noticed by the sound of gas leaking around the tube or when there is a fall in tidal volume and end-tidal carbon dioxide (CO2). Confirmation and repositioning of the ETT require direct laryngoscopy which has disadvantages such as stress response [2] and is difficult to perform intra-operatively, particularly if the patient is not in supine position. We report a case of ETT cuff herniation from glottis which was detected with the use of ultrasound.


   Case Report Top


A 28-year-old male, weighing 60 kg, of American Society of Anesthesiologists Class 1 and Mallampatti Grade 1, was posted for elective laparoscopic appendectomy. General anaesthesia was induced and tracheal intubation was performed with 8 mm ID size Portex ® endotracheal tube, the cuff was inflated with 5 ml air using minimal occlusion method and the ETT was fixed at 20 cm (at the lower lip) after confirming position by auscultation and capnography. ETT cuff pressure of 20 cm of H2O was recorded. Anaesthesia was maintained with isoflurane, N2O and O2 and controlled ventilation with the circle system. Surgery was commenced, peritoneal insufflation with CO2 was done, and the patient was then positioned in the Trendelenburg position with the right side up. After positioning the patient, partial collapse of the pilot balloon was noticed, but there was no audible leak. Cuff pressure measured was 10 cm H2O. We inflated the cuff again with 5 ml of air, but the cuff pressure remained at 10 cm H2O. The pilot balloon remained partially collapsed, and there was no leak. Despite proper fixation of the ETT, we noticed that the tube was dragged to 19 cm at the lip level, but there was no change in the SpO2 (100%) and end-tidal CO2 (30 mmHg). No leak was heard over the trachea on auscultation. Suspecting ETT cuff damage or herniation, ultrasound examination of the neck was done using a curvilinear probe (2–5 MHz) of M-Turbo ® Ultrasound Machine (Sonosite Inc., Bothell, WA, USA) in both sagittal and transverse planes. The sagittal scan of a properly placed ETT will show smooth concave hyperechoic line entering into the glottis as shown in [Figure 1]a and [Figure 1]b. This hyperechoic line is because of the ultrasound beam hitting the air inside the ETT. We refer to this line as the air-airway device (AAD) interface, similar to the air-mucosal (A-M) interface described in ultrasound of the airway without an airway device.[3] In contrast to these findings, we found a convex the AAD interface of ETT cuff just above the glottic opening [Figure 1]c and [Figure 1]d. Similarly, the transverse scan (at infrahyoid level) with a properly placed ETT will show a small convex AAD interface (ETT in short axis) below the epiglottis toward the right side (if the tube is fixed to the right angle of mouth) as shown in [Figure 2]a and [Figure 2]b. In our case, a large AAD interface, i.e., the ETT cuff was found to be compressing the epiglottis as shown in [Figure 2]c and [Figure 2]d. These findings suggest that there was herniation of cuff out of the glottis, which was also confirmed by direct laryngoscopy. As the cuff was at the glottis and over inflated, adequate ventilation was possible with a minimal leak. The ETT cuff was deflated, and the ETT was advanced by 2 cm under direct laryngoscopy. ETT position was checked with bilateral equal air entry. The cuff was inflated with 5 ml of air and cuff pressure of 22 cm H2O was recorded. The ETT and cuff position were also confirmed with ultrasound.
Figure 1: Sagittal view at the glottis inlet with curvilinear probe. (a) Ultrasonography image and (b) line diagram showing correct placement of endotracheal tube with the cuff inside the trachea. (c) Ultrasonography image and (d) line diagram - - endotracheal tube with herniated cuff outside the glottis. Hollow arrows - air-airway device (AAD) interface line artefact of the endotracheal tube wall (long axis), dotted line-shadow of hyoid bone, Hollow triangles - air-airway device (AAD) interface line of the endotracheal tube cuff. TC - Thyroid cartilage, GH - Geniohyoid muscle, HB - Hyoid bone, T - Tongue. Epi - Epiglottis. Insert-probe position and orientation

Click here to view
Figure 2: Infrahyoid transverse view with the curvilinear probe. (a) Ultrasonography image and (b) line diagram showing correct placement of the endotracheal tube with the cuff inside the trachea. (c) Ultrasonography image and (d) line diagram - endotracheal tube with the cuff herniated outside the glottis. Hollow arrows - air-airway device (AAD) interface line of the endotracheal tube wall (short axis), dotted line-shadow of hyoid bone. Hollow triangles - air-airway device (AAD) interface line artefact of the endotracheal tube cuff. SM - Strap muscle, PES - Pre-epiglottic space, Epi - Epiglottis. Insert-probe position and orientation

Click here to view



   Discussion Top


We report this case to describe the ultrasound image of cuff herniation through the glottis with the use of point of care ultrasound (POCUS). Ultrasound has become a universal tool for the anaesthesiologist. It has been used for various indications in the pre- and intra-operative period.[3],[4],[5] Airway ultrasound to confirm ETT position can avoid unnecessary laryngoscopy and its complications, particularly in the critical care unit or in the emergency wards, where patients are conscious and direct laryngoscopy is a challenge.

In our case, the cuff might have partially come out of the glottis during change in position of the patient; however, there was no audible leak and since the seal was maintained, there was no change in SpO2 and capnograph. The ETT cuff got over inflated after extra 5 ml of air was injected into the cuff which was seen as herniation of ETT cuff in on ultrasound.

Undiagnosed overinflated ETT cuff at the glottic opening can produce hoarseness voice, laryngitis and traumatic injury to the vocal cord. There are a few case reports about herniation of the ETT cuff inside the tube causing airway obstruction.[6],[7] There are various factors which can cause ETT cuff herniation such as manufacturing defect, poor tube material, nitrous oxide diffusion into the cuff and multiple uses of the tube. However, the most common cause for herniation of ETT cuff is overinflation of cuff.[8]


   Conclusion Top


In case of suspected ETT cuff herniation, airway ultrasound can be used to confirm the diagnosis.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Wright D, Baruch M. Herniation of tracheal tube cuffs: A simple teaching model. Anaesthesia 2001;56:277.  Back to cited text no. 1
    
2.
Mort TC. Emergency tracheal intubation: Complications associated with repeated laryngoscopic attempts. Anesth Analg 2004;99:607-13.  Back to cited text no. 2
    
3.
Bajracharya GR, Truong AT, Truong DT, Cata JP. Ultrasound-assisted evaluation of the airway in clinical anesthesia practice: Past, present and future. Int J Anesthesiol Pain Med 2015;1:1-2.  Back to cited text no. 3
    
4.
Isono S, Sandberg WS, Jiang Y. Do you believe what you see or what you hear? Ultrasound versus stethoscope for perioperative clinicians. Anesthesiology 2016;124:989-91.  Back to cited text no. 4
    
5.
Hatfield A, Bodenham A. Ultrasound: An emerging role in anaesthesia and intensive care. Br J Anaesth 1999;83:789-800.  Back to cited text no. 5
    
6.
Moazeni Bitgani M, Madineh H. Intraoperative atelectasis due to endotracheal tube cuff herniation: A case report. Acta Med Iran 2012;50:652-4.  Back to cited text no. 6
    
7.
Ward CF, Gamel DM, Benumof JL. Endotracheal tube cuff herniation: A cause of delayed airway obstruction. Anesth Analg 1978;57:114-6.  Back to cited text no. 7
    
8.
Hollis LJ, Almeyda JS, Mochloulis G, Patel KS. An in vitro study of tracheostomy tube cuff herniation and inflation characteristics. J Laryngol Otol 1996;110:1142-4.  Back to cited text no. 8
    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   Introduction
   Case Report
   Discussion
   Conclusion
    References
    Article Figures

 Article Access Statistics
    Viewed1462    
    Printed7    
    Emailed0    
    PDF Downloaded233    
    Comments [Add]    

Recommend this journal