Pattern of anaesthetic equipment contamination and infection prevention in anaesthesia practice at university hospitals


1 Department of Medical Microbiology and Immunology, Faculty of Medicine, Zagazig University, Zagazig, Egypt
2 Department of Medical Microbiology and Immunology, Faculty of Medicine, Fayoum University, Fayoum, Egypt
3 Department of Anaesthesia, Faculty of Medicine, Fayoum University, Fayoum, Egypt

Correspondence Address:
Dr. Rehab M Elsaid Tash
Department of Medical Microbiology and Immunology, Faculty of Medicine, Zagazig University, Zagazig, Al Sharqia Governorate
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ija.IJA_41_18

Rights and Permissions

Background and Aims: Infection control is essential in anaesthetic practice for both personnel and equipment used. This study aims to evaluate knowledge of anaesthesiologists about infection control practices and to detect the pattern of anaesthetic devices contamination. Methods: Cross-sectional observational study at two university hospitals was done. Self-administered questionnaires were distributed to 80 anaesthesiologists and 90 nursing staff. Forty-four samples were taken from rigid laryngoscopes (22 pairs from handle and blade) for detection of bacterial or fungal contamination. Same laryngoscopes were tested for occult blood. Results: The response rate among the physicians was 72% while for nurses 94.4%. The responses were variable reflecting lack of adequate knowledge and unsatisfactory compliance to infection control practices. Tested samples showed no fungal growth. Fourteen (31.8%) samples were negative for bacteriological contamination and 5/44 (11.4%) showed gram-positive bacilli; gram-positive cocci were isolated from 12 samples (27.3%) where Staphylococcus epidermidis and Staphylococcus aureus, respectively, shared 18.2% and 9.1% of the total samples. Gram-negative bacilli were isolated from 13 samples (29.5%), of which Klebsiella spp. were most frequent (11.4%). Both Pseudomonas aeruginosa and Acinetobacter baumannii were isolated from 6.8% each. Citerobacter spp. was isolated from 4.5%. Occult blood was found in 45.5% of samples. Conclusion: The current study showed contamination of ready-to-use laryngoscopes in operative theatres and ICUs.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed847    
    Printed7    
    Emailed0    
    PDF Downloaded156    
    Comments [Add]    

Recommend this journal