C-MAC® D-BLADE for awake oro-tracheal intubation with minimal mouth opening – A safe alternative to fibreoptic bronchoscope


Department of Anaesthesiology, Pain Medicine and Critical Care, All India Institute of Medical Sciences, New Delhi, India

Date of Web Publication2-Nov-2018

Correspondence Address:
Dr. Kanil R Kumar
Department of Anaesthesiology, Pain Medicine and Critical Care, All India Institute of Medical Sciences, Ansari Nagar, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ija.IJA_431_18

Rights and Permissions

How to cite this article:
Kumar KR, Sinha R, Mandal P, Chowdhury AR. C-MAC® D-BLADE for awake oro-tracheal intubation with minimal mouth opening – A safe alternative to fibreoptic bronchoscope. Indian J Anaesth 2018;62:916-8

How to cite this URL:
Kumar KR, Sinha R, Mandal P, Chowdhury AR. C-MAC® D-BLADE for awake oro-tracheal intubation with minimal mouth opening – A safe alternative to fibreoptic bronchoscope. Indian J Anaesth [serial online] 2018 [cited 2018 Dec 10];62:916-8. Available from: http://www.ijaweb.org/text.asp?2018/62/11/916/244844



Sir,

Awake fibreoptic bronchoscopy (FOB) guided intubation is considered as the technique of choice for an anticipated difficult airway. However, video laryngoscopes are gaining popularity as first line devices in these scenarios.[1] Being similar to conventional direct laryngoscope with a short learning curve, C-MAC® video laryngoscope (Karl Storz, Tuttlingen, Germany) has got rapid acceptance among anaesthesiologists. Here, we describe the role of the angulated C-MAC® D-Blade for awake orotracheal intubation in restricted mouth opening situation.

A 28-year-old male with retinal haemorrhage was scheduled for vitreo-retinal surgery under general anaesthesia. The patient had a road traffic accident 10 days back. On examination, he had mouth opening of just 1.2 cm and the trismus was not due to pain [Figure 1]a. The computed tomography scan showed fractures of anterior cranial fossa, medial orbital wall and floor. Also, there were bilateral maxillary, nasal and left zygomatic bone fractures. There were no fractures of the mandible or dentition.
Figure 1: (a) Patient with mouth opening of 1.2 cm. (b) C-MAC® D-Blade with a preshaped 7 size ETT well within the thickness of the blade

Click here to view


Our initial plan was awake orotracheal intubation with C-MAC® D-Blade. Senior anaesthesiologist experienced in performing FOB was present and difficult airway cart was kept ready. In the preanaesthesia room, the patient received inj. Glycopyrrolate 0.2 mg intramuscularly and oral cavity was anaesthetised by gargling with 2% lignocaine viscous. Inside the operation theatre, supplementary oxygen was provided through nasal prongs and inj. Fentanyl 50 mcg was given intravenously. Further airway anaesthesia was achieved with bilateral superior laryngeal nerve block using 2 ml of 2% lignocaine on each side and a transtracheal injection with 4 ml of 2% lignocaine. A 7-mm internal diameter cuffed endotracheal tube (ETT) with stylet was preshaped to the curvature of C-MAC® D-Blade.[2] The blade was inserted along the midline of the tongue till the vallecula, and the glottic aperture was visible (percentage of glottic opening score 60%) without much pressure. The preshaped ETT was passed along the curvature of the blade and glided smoothly into the glottis. Throughout the procedure the patient was comfortable and there was no coughing or gag response.

The American Society of Anaesthesiologists difficult airway algorithm advises considering video laryngoscopy as an initial approach for difficult intubation.[1] In our case, laryngeal mask airway placement was difficult with mouth opening of 1.2 cm. Likewise, FOB-guided nasal intubation was contraindicated, due to comminuted nasal and anterior cranial fossa fractures. The oral route for FOB is relatively difficult due to the tendency to go off midline and acute oral to laryngeal axis. This is especially true in restricted mouth opening patients where inserting an intubating oral airway is not possible.

We preferred D-Blade for its feasibility to use in a restricted mouth opening of 1.2 cm as it is slim and elliptically tapered.[3] D-Blade has space to accommodate ETT along its dorsal side. We chose 7 size cuffed ETT (9.6-mm outer diameter) so that it could be accommodated well within the dorsum of blade [Figure 1]b. The maximum thickness of D-Blade with the ETT alongside was just 1 cm. Furthermore, the D-Blade is hemi-moon shaped with a 40° curvature and 80° field of vision, making glottic visualisation easy without the need of undue pressure or lifting force.[4] We did not face any difficulty in gliding a preformed tube over the curvature of the blade in restricted mouth opening. But if any difficulty arises, the tube can be inserted with its curvature facing right and then rotated (90° anticlockwise) to align along the scope.[2] Along with the pertinent features of D-Blade, airway blocks made awake oral intubation smooth and tolerable.

This article highlights the feasibility of awake orotracheal intubation with D-Blade in a restricted mouth opening situation as a safe alternative to FOB.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Apfelbaum JL, Hagberg CA, Caplan RA, Blitt CD, Connis RT, Nickinovich DG, et al. Practice guidelines for management of the difficult airway: An updated report by the American Society of Anesthesiologists Task Force on management of the difficult airway. Anesthesiology 2013;118:251-70.  Back to cited text no. 1
    
2.
Shah SB, Hariharan U, Bhargava AK. C Mac D blade: Clinical tips and tricks. Trends Anaesth Crit Care. 2016;6:6-10.  Back to cited text no. 2
    
3.
Kramer A, Müller D, Pförtner R, Mohr C, Groeben H. Fibreoptic vs videolaryngoscopic (C-MAC D-BLADE) nasal awake intubation under local anaesthesia. Anaesthesia. 2015;70:400-6.  Back to cited text no. 3
    
4.
Cavus E, Neumann T, Doerges V, Moeller T, Scharf E, Wagner K, et al. First clinical evaluation of the C-MAC D-Blade videolaryngoscope during routine and difficult intubation. Anesth Analg 2011;112:382-5.  Back to cited text no. 4
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed279    
    Printed2    
    Emailed0    
    PDF Downloaded119    
    Comments [Add]    

Recommend this journal