Looped suction catheter in an i-gel™; something to worry about or much ado about nothing?


1 Department of Anaesthesia and Critical Care, Jawaharlal Institute of Postgraduate Medical Education and Research, Puducherry, India
2 Department of Anaesthesia and Intensive Care, Postgraduate Institute of Medical Education and Research, Chandigarh, India

Date of Web Publication13-May-2019

Correspondence Address:
Dr. Vikas Saini
Department of Anaesthesia and Intensive Care, Postgraduate Institute of Medical Education and Research, Sector 12, Chandigarh - 160 012
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ija.IJA_804_18

Rights and Permissions

How to cite this article:
Naik B N, Ganesh V, Saini V, Samra T. Looped suction catheter in an i-gel™; something to worry about or much ado about nothing?. Indian J Anaesth 2019;63:420-1

How to cite this URL:
Naik B N, Ganesh V, Saini V, Samra T. Looped suction catheter in an i-gel™; something to worry about or much ado about nothing?. Indian J Anaesth [serial online] 2019 [cited 2019 Oct 22];63:420-1. Available from: http://www.ijaweb.org/text.asp?2019/63/5/420/258070



Sir,

We describe a case in which a suction catheter inserted through the gastric drain channel of an i-gel™ (size 3) was found emerging into the shaft (airway channel). The semitransparent nature of the device enabled its recognition. We discuss the clinical implications of the suction catheter malpositioning in the airway device and suggest preventive and remedial measures.

A 22-year-old female patient with American Society of Anesthesiologists physical status I (ASA PS I) scheduled for nanofat grafting for a scar over left forehead was administered general anaesthesia and the airway was secured with i-gel™ (size 3). A 10 Fr suction catheter was inserted smoothly through gastric drain channel without resistance. Anaesthesia was maintained with spontaneous ventilation under a 40:60 oxygen-nitrous mixture with a total of 1 MAC sevoflurane. The surgical duration was 1 h and no issues with ventilation or oxygenation were noticed on the standard monitors. A careful examination through the semitransparent shaft of the airway device revealed a looped suction catheter which was also confirmed after removal of the i-gel at the end of surgery [Figure 1]a and [Figure 1]b.
Figure 1: (a) Intraoperative photograph of the i-gel with the retrograde suction catheter tip (Red arrowhead) seen through the semi-transparent ventilating port. (b) The course of the suction catheter after removing the i-gel en masse

Click here to view


Impaction of suction catheter in preformed nasal endotracheal tube (ETT) has been reported in a paediatric dental patient.[1] Resistance to pull out suction catheter, which has entered Murphy eye of ETT has been described by Jagannathan and Pak.[2] Gupta et al. reported a case of impaction of suction catheter within ETT even though the catheter was of recommended size and there was no knotting inside the ETT.[3] Dubey and Sanjeev have reported impaction of suction catheter in paediatric ETT as these ETT have a smaller lumen and sharper curve.[4] Perilli et al. proclaimed improved success rate during ProSeal laryngeal mask airway (PLMA) insertion using suction catheter. They showed the main cause of failed PLMA insertion with suction catheter technique was misplacement of the suction catheter.[5] Our patient did not have any gastric regurgitation but a looped suction catheter increases risk of aspiration of gastric contents because i-gel has lower sealing pressures among the second generation supraglottic airway devices, making it less safe for patients at increased risk of aspiration. This report brings to light that such unexpected complications may occur and thus a prompt visual inspection of the transparent body of the i-gel after insertion should be made to recognize any malpositioning of the suction catheter. Correct placement of the suction catheter in the gastric tube should be assessed by conventional auscultation method. Supraglottic airway devices are associated with various complications.[6] This is the first report of a malpositioned suction catheter within the i-gel.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Takrouri MS, Nafakh R, Abbas AA. Suction catheter impaction in preformed nasal endotracheal tube (PNETT) during pediatric dental anesthesia-hazard notice. Internet J Anesthesiol 2008;22:1.  Back to cited text no. 1
    
2.
Jagannathan N, Pak TY. An unusual complication of endotracheal tube suctioning. J Anesth 2009;23:170-1.  Back to cited text no. 2
    
3.
Gupta A, Mohta A, Kamal G, Bathla S. Impaction of suction catheter-Complication of endotracheal suctioning. Indian J Crit Care Med 2010;14:222.  Back to cited text no. 3
    
4.
Dubey PK, Sanjeev OP. Impaction of suction catheter during paediatric anaesthesia. Anaesthesia 2013;68:102-18.  Back to cited text no. 4
    
5.
Perilli V, Aceto P, Sacco T, Martella N, Cazzato MT, Sollazzi L. Suction catheter guided insertion of ProSeal laryngeal mask airway: Experience by untrained physicians. Indian J Anaesth 2014;58:25-9.  Back to cited text no. 5
[PUBMED]  [Full text]  
6.
Michalek P, Donaldson W, Vobrubova E, Hakl M. Complications associated with the use of supraglottic airway devices in perioperative medicine. Biomed Res Int 2015;2015:746560.  Back to cited text no. 6
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed509    
    Printed8    
    Emailed0    
    PDF Downloaded86    
    Comments [Add]    

Recommend this journal