Troubleshooting ProSeal LMA


1 Senior Consultant, Department of Anaesthesiology, Pain and PerioperativeMedicine, Sir Ganga RamHospital,Old Rajinder Nagar,New Delhi-110 060, India
2 Senior Consultant, Chairperson, Department of Anaesthesiology, Pain and PerioperativeMedicine, Sir Ganga RamHospital,Old Rajinder Nagar,New Delhi-110 060, India
3 Emeritus Consultant, Department of Anaesthesiology, Pain and PerioperativeMedicine, Sir Ganga RamHospital,Old Rajinder Nagar,New Delhi-110 060, India

Correspondence Address:
Bimla Sharma
Department of Anaesthesiology, Pain and PerioperativeMedicine, Sir Ganga RamHospital, Old Rajinder Nagar, New Delhi-110060
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


PMID: 20640203

Rights and PermissionsRights and Permissions

Supraglottic devices have changed the face of the airway management. These devices have contributed in a big way in airway management especially, in the difficult airway scenario significantly decreasing the pharyngolaryngeal morbidity. There is a plethora of these devices, which has been well matched by their wider acceptance in clinical practice. ProSeal laryngeal mask airway (PLMA) is one such frequently used device employed for spontaneous as well as controlled ventilation. However, the use of PLMAat tunes maybe associated with certain problems. Some of the problems related with its use are unique while others are akin to the classic laryngeal mask airway (eLMA). However, expertise is needed for its safe and judicious use, correct placement, recognition and management of its various malpositions and complications. The present article describes the tests employed for proper confirmation of placementto assess the ventilatooy and the drain tube functions of the mask, diagnosis of various malpositions and the management of these aspects. All these areas have been highlighted under the heading of troubleshooting PLMA. Many problems can be solved by proper patient and procedure selection, maintaining adequate depth of anaesthesia, diagnosis and management of malpositions. Proper fixation of the device and monitoring cuff pressure intraopera­tively may bring down the incidence of airway morbidity.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3544    
    Printed95    
    Emailed1    
    PDF Downloaded761    
    Comments [Add]    

Recommend this journal