• Users Online: 1306
  • Print this page
  • Email this page


 
LETTER TO EDITOR
Year : 2010  |  Volume : 54  |  Issue : 6  |  Page : 587-588 Table of Contents     

Guide wire loss during central venous cannulation


Department of Anaesthesiology and Critical Care, GB Pant Hospital, New Delhi, India

Date of Web Publication22-Nov-2010

Correspondence Address:
Pragati Ganjoo
Department of Anaesthesiology and Critical Care, GB Pant Hospital, Jawahar Lal Nehru Marg, New Delhi 110002
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.72665

Rights and Permissions

How to cite this article:
Wadehra A, Ganjoo P, Tandon MS. Guide wire loss during central venous cannulation. Indian J Anaesth 2010;54:587-8

How to cite this URL:
Wadehra A, Ganjoo P, Tandon MS. Guide wire loss during central venous cannulation. Indian J Anaesth [serial online] 2010 [cited 2020 Oct 25];54:587-8. Available from: https://www.ijaweb.org/text.asp?2010/54/6/587/72665

Sir,

Central venous (CV) catheterisation is a routine procedure in intensive care units (ICU), with an overall complication rate of 12%. Loss of a complete guide wire into the circulation is one of its rare and preventable complications. [1] A 65-year-old male patient in our neurosurgical ICU required CV access on the second postoperative day, following an intracranial tuberculoma excision. CV catheterisation of the right femoral vein with the Seldinger technique was attempted in him by a first year resident doctor without supervision. During cannulation, an inadvertent leg movement by the drowsy patient caused the guide wire to slip out of the resident's fingers, into the circulation. On radiography, the guide wire was seen to have ascended in the inferior vena cava and entered the heart [Figure 1] and [Figure 2]. The migrated wire did not produce any cardiac manifestations. Heparinisation of the patient for guide wire extraction was deferred for 2 days due to the recent neurosurgery. The guide wire was then removed percutaneously using a gooseneck snare under fluoroscopic guidance in the cardiac catheterisation laboratory.
Figure 1: Migrated guide wire ascending in the inferior vena cava

Click here to view
Figure 2: Intracardiac migration of the lost guide wire

Click here to view


Migration of a guide wire into the circulation can occur from any of the usual CV catheter insertion sites. [1],[2],[3] A complete guide wire may not necessarily produce any symptoms and its loss may remain unnoticed for long. [2] However, intravascular migration of a broken guide wire has the potential of causing adverse effects like vascular damage, thrombosis, embolism and arrhythmias; embolism from guide wire fragments can be fatal in up to 20% instances. [1] Cardiac tamponade manifesting 3 years after a guide wire loss has been reported as a late complication, highlighting the importance of wire extraction as soon as a diagnosis is made. [3] Retrieval is usually done by interventional radiology using gooseneck snares, endovascular retrieval forceps or Dormia baskets; surgical removal is also reported. [4]

The usual attributes for a guide wire loss are operator inexperience, inattention and inadequate supervision during catheterisation. [1] In our patient too, these were the main causes, compounded by the sudden movement of our disoriented patient. Expert operator skills and compliance with the catheterisation protocol are mandatory to prevent this complication. Firmly holding on to the tip of the guide wire at all times during catheterisation is the mainstay of prevention. [1] Prior sedation of disoriented patients may help achieve a smoother cannulation though sedatives need to be used cautiously in neurosurgical patients.

 
   References Top

1.Schummer W, Schummer C, Gaser E, Bartunek R. Loss of the guide wire: mishap or blunder? BJA, 2002;88:144-6.   Back to cited text no. 1
    
2.Guo H, Peng F, Ueda T. Loss of the guide wire: a case report Circ J 2006;70:1520-2.  Back to cited text no. 2
    
3.Pιrez-Dνez D, Salgado-Fernαndez J, Vαzquez-Gonzαlez N, Calviρo-Santos R, Vαzquez-Rodrνguez JM, Aldama-Lσpez G, et al. Percutaneous retrieval of a lost guidewire that caused cardiac tamponade. Circulation 2007;115:629-31.  Back to cited text no. 3
    
4.Kumar S, Eapen S, Vaid VN, Bhagwat AR. Lost guide wire during central venous cannulation and its surgical retrieval. Indian J Surg 2006;68:33-4.  Back to cited text no. 4
    


    Figures

  [Figure 1], [Figure 2]


This article has been cited by
1 Fio-guia perdido – Lições aprendidas
Daniel Rodrigues Alves,Cláudia Carvalho
Brazilian Journal of Anesthesiology. 2016;
[Pubmed] | [DOI]
2 Lost guide wire – lessons learned
Daniel Rodrigues Alves,Cláudia Carvalho
Brazilian Journal of Anesthesiology (English Edition). 2016;
[Pubmed] | [DOI]
3 Missed Central Venous Guide Wires
Krishna Pokharel,Binay Kumar Biswas,Mukesh Tripathi,Asish Subedi
Critical Care Medicine. 2015; 43(8): 1745
[Pubmed] | [DOI]
4 Accidental guide-wire loss during central venous catheterization: A report of two life-threatening cases
Ghatak, T. and Azim, A. and Baronia, A.K. and Ghatak, N.K.
Indian Journal of Critical Care Medicine. 2013; 17(1): 53-54
[Pubmed]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed1946    
    Printed107    
    Emailed0    
    PDF Downloaded331    
    Comments [Add]    
    Cited by others 4    

Recommend this journal